Higher concentration than expected
Reconstitution volume (for kits with lyophilized calibrators or controls)
The lyophilized calibrators have been reconstituted with a larger volume than stated in the Directions for Use.
Reconstitute with recommended volume of redistilled water. Mix well before using.
Dilution
The samples and/or controls have not been diluted enough.
Always use the volumes recommended for samples, calibrators, controls and enzyme conjugate solution. Check that pipettes are calibrated.